Mechanical parameters
|
Dimensions (mm)
|
95×68.5×36
|
Weight (g)
|
210
|
Detector specification
|
Detector
|
HAMAMATSU S9840
|
Wavelength Response Range (nm)
|
200-1000
|
Active Pixels
|
2048X14
|
Pixel size (µm)
|
14×14
|
Pixel well capacity (ke-)
|
130
|
Saturation Exposure(lx·s)
|
0.004
|
Saturation Output Voltage
|
0.8V
|
A/D Resolution
|
16 bit
|
Dark Current(pA/cm2)
|
40(MAX 120)
|
Photo response Non-uniformity
|
±3%
|
Sensitivity (uV/e-)
|
4.0
|
Optical Bench
|
Optical System
|
F/4, symmetrical crossed Czerny-Turner
|
Focal length (mm)
|
98.5 (input); 98.5 (output)
|
Entrance Aperture (µm)
|
5, 10, 15, 20, 50 .etc width slits or fiber (customizable)
|
Grating
|
Various kinds of scoring grating or holographic grating with different grating constants and flare wavelengths (customizable)
|
Cylindrical Lens
|
optional
|
Optical Filter
|
Band pass or long pass filters to eliminate secondary spectrum
|
Fiber and Fiber Connection
|
NA 0.22, SMA905 (customizable)
|
Spectroscopic
|
Wavelength Range
|
Depends on optical grating
|
Optical Resolution (nm)
|
0.14-10 FWHM (Depends on grating constant, width of slit and detector model)
|
SNR
|
450:1 (Full signal)
|
Wavelength Tolerance (nm)
|
About ±0.2
|
Dark noise (RMS counts)
|
50
|
Integration Time
|
10um-7.8s
|
Dynamic Range
|
2.0×108(system); 5200:1 (single acquisition)
|
Stray light
|
<0.05% at 550nm
|
Electronics
|
Power Consumption
|
400 mA×5 V DC
|
Data Transfer Speed
|
Full spectrum into memory every 2.4 ms at USB 2.0
|
Computer
|
Operating System
|
Windows /XP/Vista/7/8 (32 and 64)
|
Interface
|
USB 2.0 (480Mbps), compatible with USB 1.1
|
Pre-defined System Components
|
Name
|
Customizable Parts
|
Remark
|
Slit
|
Width (µm)
|
10, 20, 50, 100 .etc
|
Grating
|
Lines Per Unit (L/mm)
|
300, 600, 1200, 1800
|
|
Wavelength Range(nm)
|
200-1100 (User defied, about 650 range with 600 lines/mm grating)
|
Fiber
|
Core Diameter (µm)
|
9, 50, 200, 400, 600 .etc
|
Wavelength Range(nm)
|
200-1100
|
Outer Protection Layer
|
Normal or enhanced jacketing
|
Interface type
|
SMA 905 (standard) or FC, ST .etc
|
Detector
|
Model
|
Sony ILX511
|
Cylindrical Lens
|
Optional
|
Enhance system’s light gathering ability
|
Optional components
|
Name
|
Remark
|
Light Source
|
Used for wide range spectral measurements or wavelength calibration: for example halogen tungsten, xenon or argon light sources
|
Integration Sphere
|
Used to acquire uniform light
|
Reflection Measurement Accessory
|
Components used for reflection spectrum measurements: for example standard reflection board or specialized clamp
|
Transmission Measurement Accessory
|
Components used for transmission spectrum measurement: for example sample holder and various kinds of cuvettes
|
Packing List
|
Name
|
Amount
|
ToupTek Microspectrometer
|
MSPM09840A
|
Fiber with SMA 905 Interface
|
1
|
Mini USB Data Transfer Wire
|
1
|
CD of Software and SDK
|
1
|